An endoscopy purchase should include a plan for what happens after the examination. The scope, cleaning tools, disinfectant, water supply, drying equipment, storage, and staff training all need to work together. A useful buying question is: Can our team return this instrument to a documented, patient-ready condition after every procedure?
Reprocessing is the complete process of cleaning an instrument and then disinfecting or sterilizing it for reuse. Inadequate cleaning can leave biological material behind, compromising the next stage of processing. That is why a disinfectant purchase alone does not establish a reprocessing program. [1]
For compatible flexible veterinary endoscopes used on mucous membranes, we recommend ortho-phthalaldehyde, or OPA, as the preferred high level disinfectant within a complete reprocessing protocol. The process must also account for the patients being examined, the instrument’s design, and the conditions in which the veterinary team works.
What veterinary medicine can learn from human endoscopy
Human endoscopy provides a useful technical foundation. CDC’s Healthcare Infection Control Practices Advisory Committee organizes flexible endoscope reprocessing into seven essential steps: precleaning, leak testing, manual cleaning, visual inspection, disinfection or sterilization, storage, and documentation. Rinsing and drying are essential activities within that sequence. [2]
Human guidance pairs the technical steps with documented staff competency, separation of contaminated and clean work, and model-specific instructions. We recommend those same operating disciplines for veterinary teams, adapted to the practice’s caseload, facilities, and applicable requirements. [3]
A small animal hospital and an ambulatory equine practice will implement the workflow differently. Each still needs an achievable process, trained personnel, and clear criteria for releasing a scope for its next patient. Our recommendation is to make complete reprocessing the routine after every patient, without relying on the apparent absence of infectious disease to justify fewer steps.
High level disinfection and sterilization are different endpoints
The intended use of an instrument determines its processing category. The distinction is the same in veterinary and human medicine. [4,5]
| Instrument use | Processing principle |
| Contact with mucous membranes, as in routine flexible gastrointestinal or respiratory endoscopy | Thorough cleaning followed by at least high level disinfection, using a compatible process. |
| Entry into normally sterile tissue or the vascular system | Cleaning followed by an appropriate, validated sterilization process. |
| Contact only with intact skin or environmental surfaces | Cleaning and the appropriate level of surface disinfection; these products are not interchangeable with endoscope disinfectants. |
High level disinfection provides extensive microbial inactivation but does not reliably eliminate all bacterial spores. Sterilization is a distinct process intended to eliminate all viable microorganisms, including spores. An endoscope processed with OPA should be described as high level disinfected. The term sterile is reserved for an instrument that has undergone an appropriate, validated sterilization process and has been protected from recontamination. [4,6]
OPA provides the level of disinfection appropriate for many routine flexible endoscopy procedures in both veterinary and human medicine. Procedures entering sterile tissue need a different processing plan. Reusable biopsy forceps and other accessories that breach the mucosal barrier also require sterilization; they should not simply be assigned the scope’s OPA disinfection protocol. [7]
Small animal and large animal medicine have different risk profiles
Small animal practice
The infection control plan for dogs and cats needs to account for enteric and respiratory disease, multidrug-resistant bacteria, and patients whose underlying illness makes infection more consequential. Veterinary infection prevention guidance specifically addresses managing patients with suspected gastrointestinal, respiratory, and resistant bacterial infections. [5]
Nonenveloped viruses, including parvoviruses, are an additional reason to examine a disinfectant’s actual antimicrobial claims. AAHA’s guidance distinguishes the activity of chemical classes and emphasizes that formulation and concentration affect performance. A product suitable for routine environmental cleaning may lack the necessary activity for a particular exposure. [8]
For endoscopy buyers, the practical questions extend beyond chemistry. Can staff access and process every channel in the slim scope being purchased? Are the necessary brushes and adapters readily available? Can the hospital maintain clean and contaminated work areas while several patients are being treated?
Equine and other large animal practice
In equine medicine, strangles adds a particularly important endoscopy-related concern. Clinically normal horses can carry Streptococcus equi, contaminated equipment can transmit infection, and guttural pouch endoscopy and lavage are central to carrier assessment. The AAEP’s revised 2026 guidance calls for disinfecting equipment between horses when sampling multiple animals on a farm. [9]
Salmonella presents another kind of challenge. Horses may shed it without obvious clinical disease, and contaminated equipment, hands, clothing, and footwear can contribute to transmission. It also has zoonotic implications. Scope processing therefore belongs within the practice’s broader patient isolation and environmental biosecurity program. [10]
The buying implications are concrete: long scopes need suitable basins and storage; mobile teams need protected transport, adequate water, and a defined place to complete processing. A portable endoscopy system should be purchased with a plan for handling the contaminated instrument before the team moves to its next appointment.
Large animal practice also extends beyond horses. When an exposure involves unusually resistant organisms, seek an organism-specific, device-compatible plan. Laboratory research has demonstrated the resistance of Cryptosporidium oocysts to several disinfectants, including OPA. The investigators nevertheless concluded that complete cleaning, high level disinfection, and drying could prevent endoscope-related transmission. Evaluate the entire process when assessing these exposures. [11]
Strangles testing also depends on avoiding diagnostic carryover
Reprocessing can affect sample interpretation as well as infection control. A 2023 study of experimentally contaminated endoscopes found no positive postprocessing cultures, but residual S. equi DNA was frequently detected by qPCR. The OPA group had more positive qPCR results than the accelerated hydrogen peroxide group. The study used two scopes and a protocol with acknowledged limitations, including a final rinse that differed from the OPA manufacturer’s instructions. It does not establish that every labeled OPA process produces the same outcome. [12]
A separate study that included manual protocols and an automated reprocessor also found that no tested protocol completely eliminated DNA detection after experimental contamination. [13]
The practical lesson is to treat sample carryover as a separate quality concern. Unexpected results warrant review of cleaning, rinsing, collection technique, and sampling controls with the diagnostic laboratory. A positive PCR result in a horse must not simply be dismissed as dead DNA; AAEP considers PCR-positive, culture-negative horses potentially contagious. [9]
The seven essential steps
The following framework supports equipment selection and staff training. Build the working procedure from the current instructions for use, or IFUs, for the exact scope, accessories, chemicals, and processing equipment. Resolve conflicting instructions with the manufacturers before use. [2]
1 Preclean immediately at the point of use
Begin while the scope is still at the procedure location. Remove gross contamination and perform the prescribed channel flushing or aspiration before material dries. Precleaning is followed by the full cleaning process; it does not make the instrument ready for another patient. [2]
Transport the contaminated scope in a closed, protective container appropriate for the instrument. Human guidance recommends containment that limits exposure and environmental contamination while keeping surfaces moist where the device instructions permit. [14]
For veterinary teams, assign this task before the case begins. The person monitoring recovery may need another trained team member to take responsibility for the scope. Field teams should bring the precleaning supplies and transport container to the examination location.
If processing is delayed beyond the scope manufacturer’s specified window, use its delayed-processing instructions. There is no universal waiting period that makes every instrument safe to leave untreated. [3]
Buyer question: Does the package include point-of-use cleaning supplies, a suitable transport container, and clear instructions for delayed processing?
2 Leak test before manual cleaning and immersion as directed
Leak testing checks instrument integrity before the scope is exposed to cleaning fluids. A defect can allow fluid into areas not designed to become wet. The test needs to be part of each applicable reprocessing cycle, not reserved for an instrument that already looks damaged. [7]
The sequence is model-specific. For example, current human endoscope manufacturer instructions include dry connection surfaces, checking the tester, connecting and pressurizing before immersion, and moving the bending section during the test. These details explain why a generic instruction to put the scope in water and look for bubbles is insufficient. [15]
If the scope fails, remove it from clinical use and follow the manufacturer’s failed-test and service instructions. Do not improvise a prolonged soak to finish the day’s work. Maintain the required protective caps and connections throughout processing. [15]
For a veterinary buyer, the tester, correct adapter, and failed-test response should be included in staff instruction. Ask who will provide guidance if a failure occurs after hours and how the contaminated instrument should be prepared for service.
Buyer question: Is the correct leak tester supplied, and can the team demonstrate the entire test and failure response?
3 Manually clean every required surface and channel
Manual cleaning physically removes material that could interfere with disinfection. Use a compatible detergent or enzymatic cleaner, prepared at its specified dilution and temperature. Remove detachable components and clean their accessible surfaces and recesses as directed. [1,3]
The scope’s channel map matters. Human manufacturer instructions distinguish the instrument and suction pathways from air/water and auxiliary-water pathways, which may need different adapters and flushing steps. Brush the channels designed for brushing, using the prescribed diameter and length; flush the channels according to their individual instructions. A brush passed through the working channel does not complete the whole instrument’s cleaning process. [16]
Clean unused channels too. Veterinary guidance identifies omission of unused instrument and suction channels as a problem that can permit contamination and confound subsequent sampling. [5]
After the prescribed detergent exposure, rinse away cleaning residues and remove excess water as directed before disinfection. Prepare fresh cleaning solution according to its instructions. [16,17] Discard disposable brushes after use; clean and reprocess reusable brushes between uses according to their instructions. [7]
Do not substitute household detergent or a stall disinfectant because it is readily available. The purchasing decision should include an ongoing supply of the correct cleaner and cleaning accessories.
Buyer question: Can the supplier show exactly how every channel is cleaned, and provide a complete list of brushes, adapters, and consumables?
4 Inspect the cleaned instrument before disinfection
Inspect the exterior, distal tip, accessible ports, removable components, and other areas specified by the manufacturer. Good lighting and appropriate magnification help reveal retained material or damage. Residual soil calls for repeat cleaning; a damaged device needs assessment before it returns to use. [2,7]
Additional inspection or residue tests may assist training and quality review. ATP or protein results cannot prove sterility or successful high level disinfection; current multisociety guidance does not establish them as predictors of reduced infection risk. [14]
For buyers, ask the supplier to demonstrate the inspection step on the actual model. The team should be able to recognize the difference between a normal seam, retained debris, and an area that needs service review. Also ask what inspection can and cannot reveal about internal channels.
Buyer question: What lighting, magnification, inspection tools, and staff training are needed to evaluate this scope after cleaning?
5 Disinfect or sterilize then rinse and dry completely
For a compatible flexible scope used on mucous membranes, this is where OPA high level disinfection belongs. Instruments that require sterilization need their validated sterilization pathway instead.
Control the chemical exposure. Fully immerse the immersible scope and components as directed, and fill and perfuse all required channels so trapped air cannot prevent disinfectant contact. Maintain the full labeled exposure time and temperature. [17,21]
Check concentration and reuse life separately. For manual processing, Metrex specifies up to 30 days of reuse for MetriCide OPA Plus. [19] Use the dedicated MetriCide OPA Plus Solution Test Strips before each use to verify its 0.3% minimum recommended concentration. [25] Reuse limits and test requirements belong to the specific formulation and processing method. Replace solution that fails its concentration test, and do not extend the maximum reuse period because a strip still passes. [19,25]
OPA products have different instructions. The following U.S. manual-use examples illustrate why the exact product must be identified in the practice’s written protocol. [6,18,19,24]
| Product example | Manual HLD exposure | Maximum manual reuse period |
| MetriCide OPA Plus | 12 minutes at 20°C | Up to 30 days, subject to label conditions. |
| RAPICIDE OPA/28 | 10 minutes at 20°C | Up to 28 days, subject to label conditions. |
| CIDEX OPA | 12 minutes at 20°C | Up to 14 days, subject to required concentration testing. |
These are chemical contact times, not total reprocessing times. Use the exact product’s current instructions for the market where it is supplied, including concentration testing, temperature, rinsing, and reuse limits.
Rinse thoroughly after disinfection. Remove disinfectant from exterior surfaces and all channels using the specified water quality, volumes, and rinse sequence. The CIDEX OPA manual workflow, for example, specifies three separate rinses with fresh water. Inadequate rinsing can expose the next patient to chemical residue. [17]
Dry before release or storage. Dry the exterior and channels using the manufacturer-directed method. Current human guidance calls for pressure-regulated instrument air or HEPA-filtered air for lumens and warns that an AER’s final air purge may not achieve complete drying. Alcohol flushing is used when directed by the applicable instructions; it does not replace adequate drying or high level disinfection. [14,20]
Buyer question: What verifies concentration, temperature, exposure, channel contact, rinsing, and complete drying, and what happens when any requirement is not met?
6 Store the scope so it remains ready for use
Storage protects the completed work. A processed scope needs a clean, dry, protected environment and a clear indication of its status. It should not return to a contaminated counter or be stored wet in its ordinary transport case. [7]
Choose a cabinet or storage system that follows the instrument’s instructions. Vertical storage should avoid tight coils and contact with the cabinet floor. A horizontal system is appropriate when specifically designed and validated for the instrument. Manage valves and caps as directed so the chosen storage arrangement supports dryness. [3]
The physical fit deserves attention before purchase. A storage cabinet that accommodates a short bronchoscope may not accommodate an equine gastroscope correctly. Ask for a demonstration with the actual scope length, control body, umbilical, and accessories. Evaluate cabinet cleaning and maintenance as part of the purchase.
Set a written policy for storage intervals, handling, and events that trigger reprocessing. A wet scope, a dropped instrument, or uncertain processing history needs a defined response. The supplier should help the practice establish a policy consistent with the equipment instructions.
Buyer question: Where will this exact instrument be stored, and how will staff identify its processing status before the next examination?
7 Document the process and maintain staff competency
Create a record that connects the patient, the specific scope, and its reprocessing. Include the operator, relevant times, leak-test and inspection results, chemical checks, disinfection or sterilization record, and release status. Record failures, corrective actions, and equipment maintenance. This makes a problem traceable to particular instruments and patients. [2]
Assign responsibility for reviewing the process. Staff need practical, model-specific training before independent work and reassessment when equipment, chemicals, instructions, or performance problems warrant it. Human guidance places competency and written policies alongside the processing steps themselves. [3]
A low-volume practice can use a well-designed paper log; a larger program may prefer electronic records. What matters is whether the record can answer a real question: Which scope was used, how was it processed, and was it cleared for use?
For equine field work, we recommend including the farm or premises as well as the patient identifier. That additional detail can make an equipment review more useful when a concern spans more than one location.
Buyer question: Can the practice retrieve a complete record for one scope and patient, and does the purchase include training for the people who will actually reprocess it?
Why we recommend OPA
OPA is a practical choice for compatible, heat-sensitive flexible endoscopes. Ready-to-use products eliminate an activation step, and their instructions define the exposure conditions for high level disinfection. These features make OPA a useful foundation for a standardized veterinary workflow. MetriCide OPA Plus is the OPA product we most often recommend for compatible endoscopes. [6,20]
Our preference is conditional on documented compatibility with the endoscope and reusable components. High level disinfection efficacy, material compatibility, and suitability for a particular procedure are separate questions. A product can have strong antimicrobial activity and still be unsuitable for a particular instrument’s materials or intended use. [4]
OPA also requires safe handling. Follow its label and safety data sheet for ventilation, personal protective equipment, exposure response, contraindications, and disposal. Avoid describing a lower-odor product as harmless. Thorough rinsing remains part of patient protection. [17]
Other chemistries have valid applications in manufacturer-approved systems. Glutaraldehyde introduces its own exposure-management considerations; peroxide and peracetic-acid products require careful attention to materials compatibility. We do not recommend substituting a general environmental disinfectant for a validated endoscope product, or assuming every formulation within a chemical family behaves identically. [20,21]
Where an automated endoscope reprocessor fits
A compatible automated endoscope reprocessor is our preferred investment for making reprocessing more consistent across a veterinary team. Human multisociety guidance favors automated high level disinfection for greater reliability and reduced staff exposure to disinfectants. [14]
For buyers, the gains to evaluate are repeatable processing, reduced hands-on work, and a workflow that is easier to supervise. Automation standardizes supported reprocessing stages and reduces reliance on staff manually carrying out every part of those stages. [21] During a demonstration, compare the staff time required, the clarity of cycle-status displays and alerts, and the records available for reviewing completed cycles. Where electronic documentation is offered, ask how a cycle is linked to the scope, operator, and patient.
These gains build on thorough manual cleaning, correct channel connections, and maintained equipment. [21,22] Evaluate the full workflow, including loading, unloading, drying, and storage, to determine how well the system supports the practice’s staffing and procedure schedule.
Rutledge Medical’s TYPHOON AER is an option to evaluate using these criteria. [23] Apply the same expectations for consistency, practical labor savings, documentation, training, and service support to any system under consideration.
What a complete reprocessing quote should include
Ask suppliers to price the whole workflow so proposals can be compared on the same basis.
| Purchase area | What to request |
| Written compatibility | Exact scope models, channel connections, cleaner, disinfectant, and supported processing method. |
| Cleaning and testing equipment | Leak tester, adapters, channel brushes, flushing devices, basins, and protected transport. |
| Facility preparation | Required water quality, temperature and supply, drainage, electrical service, ventilation, and suitable work surfaces. |
| Drying and storage | The equipment needed to dry all channels and store the actual scope lengths correctly. |
| Recurring costs | Chemistry, concentration strips, brushes, filters, connectors, PPE, maintenance, and staff time. |
| Training and records | Hands-on instruction, competency assessment, written procedures, and retrievable processing records. |
| Service continuity | Preventive maintenance, repair support, loaner arrangements, and a documented contingency plan. |
Request a demonstration that starts when the examination ends and finishes with a dry, documented, patient-ready scope. Time the staff work, the machine work, and the drying process separately. That exercise can reveal whether a proposed system supports the practice’s actual schedule and whether additional scope inventory is needed.
For a small animal hospital, test the proposed workflow against a busy procedure day. For an equine practice, test it against the realities of a farm visit and the return to the hospital. An attractive machine-cycle time has limited purchasing value if the rest of the workflow cannot be completed reliably.
The endoscopy system you choose should make the full process practical for the team who will use it. Require clear instructions, suitable accessories, training, and a demonstration of all seven steps before deciding that a quote is complete.
P.S. Three Veterinary Endoscope reprocessing Mistakes You Can Avoid
A chlorhexidine wipe-down protocol does not authorize soaking. Under our specified equine gastroscopy protocol only, chlorhexidine may be used for wiping and prescribed suction through the channels. Do not immerse the covered endoscopes in chlorhexidine gluconate; soaking can cause permanent damage. Follow the written instructions for the exact scope and protocol, and complete its required high level disinfection before reuse.A study conducted with Boehringer Ingelheim and Rutledge Medical evaluated a chlorhexidine wipe-down protocol used across more than 5,000 equine gastroscopy procedures. Rutledge Medical reports no positive PCR or culture results from the testing performed. Those findings apply to the specific protocol and assays used.
Enzymatic detergent does not disinfect an endoscope. A standard enzymatic cleaner breaks down and helps remove organic material. Cleaning must still be followed by the required disinfection or sterilization process—an enzyme soak does not accomplish both.
Rubbing alcohol does not replace high level disinfection. Where the endoscope’s instructions call for it, alcohol is used after disinfection and the required water rinse to assist drying, followed by the prescribed forced-air drying step. An alcohol wipe or channel flush alone does not complete reprocessing.
References
1. U.S. Food and Drug Administration. Reprocessing of Reusable Medical Devices.
2. CDC and HICPAC. Essential Elements of a Reprocessing Program for Flexible Endoscopes.
4. CDC. A Rational Approach to Disinfection and Sterilization. 2008 guideline; webpage updated 2023.
6. Advanced Sterilization Products. CIDEX OPA Solution, U.S. product information. Updated July 2024.
8. Stull JW, et al. 2018 AAHA Infection Control, Prevention, and Biosecurity Guidelines.
9. AAEP. Strangles Infectious Disease Guidelines. Revised 2026.
10. AAEP. Salmonellosis Infectious Disease Guidelines. Revised 2026.
15. Olympus. EVIS X1 and EVIS EXERA Endoscope Leakage Testing Guide. 2024.
16. Olympus. EVIS X1 and EVIS EXERA Endoscope Manual Cleaning Guide. 2024.
17. Advanced Sterilization Products. How to Use CIDEX OPA Solution. Wall chart AP-2000361.
18. Advanced Sterilization Products. CIDEX OPA Solution Test Strips. Updated November 2025.
19. Metrex. MetriCide OPA Plus, U.S. product information.
20. CDC. Chemical Disinfectants. 2008 guideline; webpage updated 2023.
21. CDC. Disinfection of Healthcare Equipment. 2008 guideline; webpage updated 2023.
23. Rutledge Medical. Veterinary Endoscope Reprocessor. Product information.
24. Crosstex. RAPICIDE OPA/28 High-Level Disinfectant Directions for Use. Document 50098-048 Rev C.
25. Metrex. MetriCide OPA Plus Technical Bulletin. MKT-20-1036; updated January 2021.